Maintenance Series DEMC control unit

Article number 1062

Price on request
Maintenance including calibration

Maintenance Series DEMC control unit

1. Examination of the delivered device on transport damages
2. Incoming inspection
- Setup of the device
- Test measurements with calibration dust and ambient air
3. Check and readjustment of the electrical components
- Calibration of high voltage modul
4. Cleaning and check of impactors
5. Check of sheath air flow circuit
- Replacement of dessicant
- Replacement of filter
- Cleaning and replacement of internal tubing (if necessary)
- Calibration of sheath air flow rate
- Leak test
6. Software
- Examination of correct installation (Are all components existing?
Are the program paths correct? Is the program version the
current one and is it exact?)
- If necessary: update of the software
7. Evaluation of the overall function and test measurements with
complete U-SMPS system
8. Packing of the device ready for dispatch
Small repairs are included
Excluded are the following spare parts, which must be ordered
separately:
- Blower for sheath air
- High voltage modul
- Touchscreen Display (and X-ray charger for DEMC X)