Maintenance
Maintenance KHG 10
Article number 708
Price on request
Maintenance including calibration Maintenance KHG 10
1. Incoming inspection
2. Disassembling of dilution system
3. Cleaning of dilution system
4. Installation of a new absolute filter (if necessary)
5. Control of valve
6. Check of dilution factor
7. Issue of a new calibration certificate
8. Check if particles are inside the dilution system by means of a
particle counter
9. Check of heating and electronics
Small repairs are included
2. Disassembling of dilution system
3. Cleaning of dilution system
4. Installation of a new absolute filter (if necessary)
5. Control of valve
6. Check of dilution factor
7. Issue of a new calibration certificate
8. Check if particles are inside the dilution system by means of a
particle counter
9. Check of heating and electronics
Small repairs are included
Related products