Maintenance

Maintenance AQ Guard

Article number 7083

Price on request
Fidas smart 100
Maintenance including calibration

Maintenance AQ Guard

1. Examination of the delivered device on transport damages
2. Incoming inspection
- Check of the LEDs
- Examine on contamination
- Test measurements with MonoDust 1500
3. Check and readjustment of the electrical components
- Check of the photomultiplier
- Lamp power supply (electricity/power)
- Signal cards (adjustment of the offset, characteristic, signal period
measurement, coincidence recognition)
- Offset measurement of the electronic of the signal aquisition
- Chek of the Integrated suction pump
- Calibration of the volume flow
4. Check of the optical system of the sensor
- Clean the optics
- Control and adjustment of the optical light paths in the sensor
5. Firmware
- Examination on perfect installation
6. Evaluation of the overall function and calibration
- Adjustment of the sensitivity with reference particles for
different measuring ranges
- Test measurements for the evaluation of the overall function
7. Issue of a calibration certificate for the aerosol spectrometer
with equipment data, date, settings, graphics of the test
measurements
8. Issue of a calibration certificate for the measurement of PM 2,5
and PM 10 with equipment data, date, settings, graphics of the
test measurements.
- Check of linearity of the performance-tested reference devices
with serial number 0111 or 0112
- Determining of the correction factor
- Validation of the linearity after adaptation of the correction factor
9 . Packing of the devices / make ready for dispatch
Small repairs are included